EasyManuals Logo

Invacare Perfecto2 IRC5PO2AW User Manual

Invacare Perfecto2 IRC5PO2AW
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
SECTION 5—TYPICAL PRODUCT PARAMETERS
Part No 1160877 13 Perfecto
2
™Series
SECTION 5—TYPICAL PRODUCT
PARAMETERS
Alternating Current
Type BF equipment
Unit running
Unit not running
Attention - Consider Accompanying Documents
DO NOT smoke
Class II, Double Insulated
Protected against dripping water
Electrical
Requirements: 230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
Rated Current Input: 1.4 A
Sound Level: 39.5 dBA Average
Altitude: All Perfecto
2
Models: Up to 8,000 ft (2438 meters) above sea
level without degradation of concentration levels. From 8,000 ft
to 13,129 (4000 meters) below 90% efficiency.
All Perfecto
2
Models while filling a HomeFill System:
Up to 6,000 ft (1828 meters) above sea level without degrada-
tion of concentration levels. From 6,000 ft (1828 meters) to
13,129 ft (4000 meters) below 90% efficiency
Oxygen Output
Concentration Levels:
NOTE:Stated
concentrationlevels
achievedafteriniti al
warmuppe riod
(approximately30
minutes).
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
IPX1

Other manuals for Invacare Perfecto2 IRC5PO2AW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 IRC5PO2AW and is the answer not in the manual?

Invacare Perfecto2 IRC5PO2AW Specifications

General IconGeneral
Flow Rate0.5 to 5 LPM
Warranty3 years limited
Operating Environment10°C to 35°C; up to 95% relative humidity
Power Requirements120 VAC, 60 Hz
Sound Level≤45 dB(A)
Dimensions23 in (H) x 13 in (W) x 12 in (D) (58.4 cm (H) x 33 cm (W) x 30.5 cm (D))

Related product manuals