SECTION 5—TYPICAL PRODUCT PARAMETERS
Part No 1160877 15 Perfecto
2
™Series
Cabinet: Impact Resistant flame-retardant plastic cabinet that conforms to UL
94-V0
No AP/APG Not suitable for use in the presence of a flammable anaesthetic
mixture
Standards and
Regulatory Listing: IRC5PO2AW/IRC5PAW/IRC5PO2VAW:
IEC/EN60601-1, A1, A2
IEC/EN60601-1-2
IEC/EN61000-3-2
IEC/EN61000-3-3,
IRC5PO2AW/IRC5PO2VAW Models Only
ISO8359
MDD 93/42/EEC, Annex I and IX
CE marked model IRC5PO2AW & IRC5PO2VAW
Electrical: No extension cords
Placement: No closer than 7.5 cm (3 in) from any wall, furniture, draperies, or
furniture to assure sufficient air flow
Avoid deep pile carpets and heaters, radiators or hot air registers
Floor location only
No confined spaces (Example: No closets)
Tubing: 2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-Proof
Tubing (DO NOT pinch)
Time of Operation: Up to 24 hours per day
Recommended
Storage and Shipping
Temperature: -29°C to 65°C (-20°F to 150°F) at 15-95% relative humidity
Environment: Smoke and soot-free
No confined spaces (Example: No closets)
Minimum Operating
Time: 30 Minutes

Other manuals for Invacare Perfecto2 IRC5PO2AW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 IRC5PO2AW and is the answer not in the manual?

Invacare Perfecto2 IRC5PO2AW Specifications

General IconGeneral
Flow Rate0.5 to 5 LPM
Warranty3 years limited
Operating Environment10°C to 35°C; up to 95% relative humidity
Power Requirements120 VAC, 60 Hz
Sound Level≤45 dB(A)
Dimensions23 in (H) x 13 in (W) x 12 in (D) (58.4 cm (H) x 33 cm (W) x 30.5 cm (D))