Calibration Commands
Copyright© Itech Electronics Co., Ltd. 34
CALibrate:CURRent:LEVel
This command is used to define the present calibration points.
Command syntax
CALibrate:CURRent:LEVel <level>
Arguments
P1|P2
Note:
P1, P2 must be calibrated orderly.
CALibrate:CURRent[:DATA]
Return the actual output current value at present calibration point of power
supply.
Command syntax
CALibration:CURRent [:DATA] <NRf>
Arguments
<NRf>
Example
CAL:CURR 3.0002A
Note: Please execute this order when power supply output is stable.
CALibrate:SAVe
This order is used to save the calibration parameters in a nonvolatile memory.
Command syntax
CALibration:SAVe
Arguments
None
CALibrate:INITital
This command is used to initialize the calibration parameters.
Command syntax
CALibration: INITital
Arguments
None