SCPI Command Description
Copyright© Itech Electronics Co., Ltd. 18
SYSTem:RWLock
This command is transmit through RS232 interface, the same function as
SYSTem:REMote, but the difference is: this command can lock “ LOCAL
“ button as well. When this command is executed, the “ Local “ button
can’t switch to local control mode.
Command syntax
SYSTem:RWLock
Arguments
None
Example:
SYST:RWL
SYSTem:COMMunicate:SELect
This command is used to select the communication interface.
Arguments
RS232|USB|GPIB|LAN
Example
SYST:COMM:SEL RS232
Query syntax
SYSTem:COMMunicate:SELect?
Returns
RS232|USB|GPIB|LAN
SYSTem:COMMunicate:GPIB:ADDRess
This command is used to set the address of GPIB interface.
Command syntax
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <NR1>
Arguments
<1-30>
Reset Value
15
Query syntax
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?