EasyManua.ls Logo

JCM DBV-500 - DBV-500 Calibration Procedures; Validation Sensor Calibration

Default Icon
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Part No. 960-000187R_Rev. A © 2016 JCM American Corporation
16
DBV-500 Banknote Validator JCM® Training Overview November, 2016
CALIBRATION
Calibration of the DBV-500 Unit needs to be performed when any of the
following conditions occur:
When removing and replacing the CPU Circuit Board
When removing or replacing any Sensors or Sensor Boards
After cleaning the DBV-500 Unit
If the Banknote Acceptance rate has decreased.
V
ALIDATION
S
ENSOR
C
ALIBRATION
To perform the DBV-500 Validation Sensor Calibration procedure:
1. Disconnect power from the DBV-500 Unit.
2. Set DIP Switch Block #1, Switch # 8 to ON.
3. Restore power to the DBV-500 Unit.
4. Connect the USB Type-A to Mini-B cable between your PC’s USB port and the
USB port located on the DBV-500 Unit.
5. Launch the
“JCM Tool Suite Standard Edition” software application.
6. Click the
Service Mode drop-down menu (Figure 5), then click Sensor Adjust-
ment.
7. Confirm that the DBV-500 Calibration Tool is running, as shown in Figure 8.
8. Click the Sensor Calibration function screen button (Figure 8 a).
The Sensor Calibration screen appears momentarily (refer to Figure 9).
NOTE: KS-091 Reference Paper is specified for use when calibrating the DBV-500.
NOTE: The DBV-500 Unit’s Front Bezel LED will flash Green.
a
Figure 8 DBV-500 Calibration Tool

Related product manuals