EasyManua.ls Logo

Julabo CF41 - 8.1. Configuration Menu

Julabo CF41
79 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cryo-Compact Circulators
35
8.1. Configuration
>Configuration
Control
Profile Start
Main menu
Configuration
>Setpoint : keypad /RS232/ext.Pt100/eprog
Autostart: off /on
OFF-Mode : pump off /pump on
Language : German /Englich
Reset : no /yes
Variable : control /serial/eprog
Unit : °C /°F
ESC
ESC
1.
Use the up/down keys to select the desired menu item.
2. Use the left/right keys to select the parameter (parameter blinks).
3. Press enter to store the new parameter.
Setpoint1: 150.00°C
IntAct : 148.53°C
Power : 80 %
Control : intern
RS232 : 150.00°C
ExtAct : 145.30°C
Power : 80 %
Control : extern
-OFF-
S xxxx
I xxxx
R
Pt100 : 150.00°C
ExtAct : 145.30°C
Power : 80 %
Control : extern
Setpoint
The Cryo-Compact Circulator provides four poss
ibilities for setpoint setting. The
selected mode is indicated in line 1 on the LCD DIALOG-DISPLAY.
Possible parameters:
keypadSetpoint setting via keypad or via the integrated programmer.
(factory setting)
RS232 or RS485 - Setpoint setting via the serial RS232/RS485 interface
through a PC or superordinated data system.
In the header of the VFD COMFORT-DISPLAY an
Rilluminates.
It indicates that remote control mode is set.
RS232 or RS485: see menu > Serial interface < page 61.
Important:
Connect the Cryo-Compact Circulator to a PC using an interface cable.
Check the interface parameters of both interfaces (on Cryo-Compact
Circulator and PC) and make sure they match.
(see 11.1. on page 71 )
ext.Pt100 Setpoint setting via the analog socket „ext. Pt100“ using an
external temperature sensor or an appropriate voltage/current source.

Table of Contents

Related product manuals