The menu >Limits< allows for setting the minimum and maximum values for all
important setting ranges and capacity variables.
SetMax: and SetMin: Maximum and minimum setpoint
Restriction for the adjustable temperature range.
Setting range: see working temperature range
The limitation of the operating temperature range effects the temperature
setting under the menu called up with the
key - Setpoint 1, 2, 3.
It is possible to adjust only working temperatures that lie within the limit range
set here.
Existing settings for Setpoint 1, 2, 3 and also for >OverTemp< and >SubTemp<
(see page 30) are automatically defered within the limit range.
HeatingMax: and CoolingMax:
Adjusted maximum heating / cooling.
Setting range: 0 to 100 % in steps of 1 %
Heating and cooling powers of the Cryo-Compact Circulator are adjustable.
100 % corresponds to the values in the technical specifications of the
equipment.
Factory setting: CoolingMax
CF31 100 % (fixed value)
CF41 0 to 100 % in steps of 1 %