Cryo-Compact Circulators
45
Examples:
Indication after successful start:
The started programmer displays the actually calculated setpoint in line
1. The value changes until the final temperature of the section is reached
(temperature ramp).
1
st
line: Setpoint of the programmer
2
nd
line: Actual temperature value
at internal control = IntAct: xxx.xx
at external control = ExtAct: xxx.xx
3
rd
line: Selected profile and the actual section (step).
4
th
line: With the keys the following information can be
displayed in line 4.
• Remaining time of the actual section (standard display)
• Remaining total time: Profile x number of repetitions
(All: Days/hours:minutes:seconds)
Particularitys:
1.
If a total time of more than 999 days is calculated, only the hours are
displayed on the unit.
(All: hours:minutes:seconds)
2. If the time period of a section is edited after start, then the remaining
total time can no longer be displayed.
• The data of a section as stored under „Edit Profile“.
Use the keys
to successively indicate the sections of a profile.
Unused sections or incorrect data are skipped.
A JULABO programmer of generation II allows editing of a running
profile. A programmer of the previous generation requested the profile to
be interrupted, edited and then started again.
Start and quit the edit mode:
• After start, press the
key to enter the submenu „Edit Profile“.
Press