Remote control
74
IN commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of Cryo-Compact Circulator
VERSION
None
Number of software version (V X.xx)
STATUS
none
Status message, error message (see page 75)
IN_PV_00
none
Actual bath temperature.
IN_PV_01
none
Heating power being used (%).
IN_PV_02
none
Temperature value registered by the external Pt100 sensor.
IN_PV_03
none
Temperature value registered by the safety sensor.
IN_PV_04 none Setpoint temperature (SafeTemp“) of the excess temperature
protection
IN_SP_00
none
Working temperature „Setpoint 1“
IN_SP_01
none
Working temperature „Setpoint 2“
IN_SP_02 none Working temperature „Setpoint 3“
IN_SP_03
none
High temperature warning limit „OverTemp“
IN_SP_04
none
Low temperature warning limit „SubTemp“
IN_SP_05 none Setpoint temperature of the external programmer
(REG+E-PROG) .
IN_SP_06
none
Temperature indicatio in °C or °F
IN_SP_07 none 1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotation speed
after start.
IN_SP_08
none
Value of a flowrate sensor connected to the E-Prog input
IN_PAR_00
none
Temperature difference between working sensor and safety sensor
IN_PAR_01 none Te - Time constant of the external bath.
IN_PAR_02
none
Si - Internal slope
IN_PAR_03
none
Ti - Time constant of the internal bath.
IN_PAR_04 none CoSpeed - Band limit (max. difference between the temperatures
in the internal bath and external system).
IN_PAR_05 none Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
IN_PAR_06
none
Xp control parameter of the internal controller.
IN_PAR_07
none
Tn control parameter of the internal controller.
IN_PAR_08 none Tv control parameter of the internal controller.
IN_PAR_09
none
Xp control parameter of the cascade controller.
IN_PAR_10
none
Proportional portion of the cascade controller.