EasyManua.ls Logo

Julabo FC1200S - Draining

Julabo FC1200S
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
22
5.7. Draining
Notice:
¾ Do not drain the bath fluid while it is hot or cold!
Check the temperature of the bath fluid prior to draining (by
switching the unit on for a short moment, for example).
¾ Store and dispose of the used bath fluid according to the
environmental protection laws.
MIN
MAX
0
1
2
3
4
5
6
1.
MIN
MAX
0
1
2
3
4
5
6
1.
9.
8.
Turn the mains switch (1.) off.
Hold the venting grid, pull out and remove.
Take the drain tubing (8.) out of the holder and hold it
into a pail.
Open the drain tap (9.) and empty the unit
completely.
Close the drain tap and replace the drain tubing into
the holder.
Replace the venting grid.
MIN
MAX
0
1
2
3
4
5
6
9.
8.
1.
In case the recirculating cooler is placed on the floor,
the unit may be drained using a suction pump unit.
Connect the drain tubing (8.) to the suction pump
unit.
Open the drain tap (9.).
Switch the pump on and fully empty the unit.

Related product manuals