EasyManua.ls Logo

Julabo FC1200S - Page 34

Julabo FC1200S
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical connections
34
Alarm output (M)
1
2
3
This potential-free change-over contact is activated in case
of an alarm.
Pins 2 and 3 are connected under the following conditions:
- alarm
- status "OFF" and "rOFF"
- mains switch "off"
Switching capacity max. 30 W / 40 VA
Switching voltage max. 125 V/
Switching current max. 1 A
Serial interface (K)
9
6
5
1
This interface is a 9-pole connector:
Pin 2 RxD Receive Data
Pin 3 TxD Transmit Data
Pin 5 Gnd
Pin 7 RTS Request to send
Pin 8 CTS Clear to send
Interface correspondence:
Circulator Computer Circulator Computer
9-pole 25-pole 9-pole 9-pole
Pin 2 RxD Pin 2 TxD Pin 2 RxD Pin 3 TxD
Pin 3 TxD Pin 3 RxD Pin 3 TxD Pin 2 RxD
Pin 5 GND Pin 7 GND Pin 5 GND Pin 5 GND
Pin 7 RTS Pin 5 CTS Pin 7 RTS Pin 8 CTS
Pin 8 CTS Pin 4 RTS Pin 8 CTS Pin 7 RTS
Accessories:
Order No. Description
8 980 073 RS232 interface cable 9-pol./9-pol. , 2,5 m
8 900 110 USB interface adapter cable