(with T-pump) FCW2500T
Working temperature range °C -25 ... 80
Cooling capacity °C
+20 +10 +5 0 -10 -20
(water-glycol) kW 2.5 2.0 1.8 1.5 0.8 0.25
Refrigerant R404A
Heater capacity kW 1.2
Pump capacity:
Pressure max. psi /bar 51/3.5
Flow rate max. Lpm 28
Pump connections M16x1
Noise level, 1 m distance dBA 70
Filling volume l 8 ... 11
Dimensions (WxLxH) in. 18x24x19
Shipping weight Lbs/kg 164/74
Mains power connection V/Hz 230/60
Total power consumption W 2800
(with S-pump) FC1200S FCW1200S
Working temperature range °C -15 ... 80 -15 ... 80
Cooling capacity °C 20 10 5 -10
(water-glycol) kW 0.12 0.85 0.65 0.26
Refrigerant R134a
Heater capacity kW 1.2
Pump capacity:
Pressure max. psi /bar 17.4/1.2
Flow rate max. l/min 22 / 15
with tubing connections mm dia. 13.5 / 9.5
Pump connections M16x1
Noise level, 1 m distance dBA 57 53
Filling volume l 8 ... 11
Dimensions (WxLxH) mm 460 x 610 x 490
Ambient temperature °C 5 ... 40
Shipping weight Lbs/kg 145/66 150/68
Mains power connection V/Hz 208-230/60 or 230/50
Total power consumption W 2800
All data have been determined at mains voltage: 230 V / 50 Hz
ambient temperature: 20 °C bath liquid: water-glycol