12.3. Status messages
Message Description
01 MANUAL START Waterbath in keypad control mode.
02 REMOTE STOP Waterbath in „OFF“ state
03 REMOTE START Waterbath in remote control mode
12.4. Error messages
Message Description
-01 TEMP / LEVEL ALARM Safety temperature or low liquid level alarm
-03 EXCESS TEMPERATURE
WARNING
High temperature warning „ „.
-04 LOW TEMPERATURE WARNING
Low temperature warning „ „.
-05 TEMPERATURE MEASUREMENT
ALARM
Error in measuring system
-07 I
2
C-BUS WRITE ERROR
-07 I
2
C-BUS READ ERROR
-07 I
2
C-BUS READ/WRITE ERROR
Internal error
-08 INVALID COMMAND
Invalid command
-10 VALUE TOO SMALL
Entered value too small
-11 VALUE TOO LARGE
Entered value too large
-12 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the adjusted range for the high
and low temperature warning limits. But value is
stored.
-13 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
Invalid command in current operating mode