Instructions for use ELECTROmatic M/C and PM/PC
2 Safety | 2.8 Electromagnetic fields
Medical electrical devices are subject to special precautions regarding electro-
magnetic compatibility and must be installed and operated in accordance with
the tables of electromagnetic compatibility.
See also:
2 12 Information about electromagnetic compatibility, Page 72
High-frequency communications devices may interfere with medical electrical
devices.
Ask patients if they have a cardiac pacemaker or other system implanted
before you start the treatment.
Comply with the tables of electromagnetic compatibility during installation
and commissioning.
If the device needs to be used in the immediate vicinity of other equipment,
monitor the device or system for malfunctions.