7
ELECTROtorque 4890/K200.
A 1.6 Disinfecting the unit
In general, use only disinfectants
which are approved for use on
painted surfaces and plastic surfaces in
dental medicine.
Use of alcoholic disinfectants
An important requirement is that the
disinfectants be used in accordance
with the manufacturer’s instructions.
Disinfectants are available on the market in
various concentrations.
As a guideline for the use of disinfectants,
the ingredients present in the preparations
tested and found to be satisfactory are
indicated below:
The values shown are maximum values
and may not be exceeded.
96% ethanol
= max. 40 g / 100 g of disinfectant
Propanol
= max. 35 g / 100 g of disinfectant
25% glutardialdehyde
= max. 75 mg / 100 g of disinfectant
Ethylhexanal
= max. 10 mg / 100 g of disinfectant
Formaldehyde solution
= max. 10 mg / 100 g of disinfectant
Glyoxal
= max. 165 mg / 100 g of disinfectant
KaVo will assume no liability if the
preparations used contain ingredients
differing from those stated by us.