Cleaning and disinfection of the handle of the S table and holder pads
▶ Press the push button and pull off the handle.
▶ Clean the handle under flowing water.
▶ Disinfect the white and grey handle by wiping or thermodisinfection (minor dis‐
colouration may occur).
▶ Thermodisinfect the holder pad.
Sterilisation of the handle of the S table and holder pads
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise the white standard handle.
▶ Do not sterilise holder pads.
The grey handle (Mat. no. 1.002.4489) can be sterilised (see symbol on the handle).
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit