Sterilisation
All sterilisable products are temperature resistant up to max. 138 °C.
KaVo recommends sterilisation in a steam steriliser (autoclave) according
to EN 13060 / ISO 17665-1, e.g. STERIclave B 2200 / 2200 P (KaVo) or
Citomat K-Series (Getinge).
▶ Sterilise with a triple fractionated vacuum (B class steriliser); hold for
4 minutes; 134°C ± 1 °C; overpressure: 2.13 bar.
Storage ▶ Rehabilitated products should be stored protected from dust with min‐
imum exposure to germs in a dry, dark and cool space.
▶ Comply with the expiration date of the sterilised item.
2.2.1 Overview ESTETICA E50 reprocessing
Reprocessing items/medical devi‐
ces
Cleaning Disinfection Sterilisation
manual automatic manual automatic automatic
Surfaces
Device
X
X
Cushion
X
X
Gripping sleeve, white
X X X X
Gripping sleeve, grey
X X X X X
Lamp and reflector
X
X
Dentist element
Syringe sleeve for 3F and MF sy‐
ringes
X
X
X
Cannula for 3F and MF syringes
X
X
X
Holder pad
X X X X
Instrument holder
X
X
Assistant element
Connection parts of the suction
hose
X X X X
Patient unit
Tumbler support
X
X
Tumbler filler
X X X X
Glass spittoon bowl
X
X
Porcelain spittoon bowl
X X X X
2.3 Reprocessing of the surfaces of treatment unit / upholstery
Position the chair for cleaning
▶ Adjust the backrest to vertical position and move the chair upward.
See also: Instructions for UseESTETICA E50
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit /
upholstery
5/32