2.6.3 Preparation of the spittoon
Cleaning and disinfection of the spittoon bowl
CAUTION
Damage caused by inappropriate cleaning agents.
▶ Do not use acid-containing, sand-containing, and form-forming cleaning
agents.
▶ Add a dose of DEKASEPTOL gel to the spittoon bowl and distribute the agent
with a brush.
▶ For cleaning, allow the gel to act for approx. 5 minutes.
▶ For disinfection, allow the gel to act for approx. 15 minutes.
▶ After letting it work, press the "Spittoon bowl" button.
▶ Clean the spittoon bowl thoroughly using a brush.
▶ For thorough cleaning or thermodisinfection, rotate the spittoon bowl counter‐
clockwise and lift it.
CAUTION
Damage caused by thermodisinfection.
▶ Do not subject spittoon bowls made of glass to thermodisinfection.
The spittoon bowl made of porcelain can be subjected to thermodisinfection.
2.7 Automatic hygiene function of the water-conducting system
Note
Intensive disinfection kit required.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.7 Automatic hygiene function of the water-conducting
system
18/32