Note
Inspect the O-ring (Mat. no. 0.200.6128) for damage and exchange if necessary to
prevent water or oxidation damage.
▶ Remove tumbler holder ① (Mat. no. 0.764.8541) by pulling upwards.
▶ Take off the tumbler holder ③ (Mat. no. 0.764.6153).
▶ Clean the tumbler holder and tumbler filler under running water.
▶ Wet the O-ring ② with silicone grease (Mat. no. 1.000.6403).
CAUTION
Damage caused by thermodisinfection.
▶ Do not subject the instrument holder to thermodisinfection.
The tumbler filler can be subjected to thermodisinfection.
Sterilisation of the tumbler holder and filler
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise the tumbler holder and filler.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit