through the supply of fresh, clear water (limited rinsing). The water is sucked out of
the Aquamat through the filter housing to the separation system. The spittoon bowl
rinsing is alternately turned on and off.
The HYDROclean function prevents the coagulation and adhesion of proteins inside
the drainage system and improves the transport of fine particles.
Requirement
Suction hoses remain hanging in the holder.
Start the HYDROclean function
▶ Press the "HYDROclean" button.
HYDROclean is in the preparation stage.
▶ Press the "HYDROclean" button.
This starts-up the HYDROclean function.
The top LED goes dark when the process is completed.
The process can be discontinued at any time by pressing the "HYDROclean" button.
2.9 Preparation of the amalgam separator
You can use a "Dürr" amalgam separator or an "external suction kit for connecting
to a central amalgam separator" or a separation kit.
Optimal care of the amalgam separation system includes the correct and regular
use of the HYDROclean function as well as the care of the suction hoses and spit‐
toon bowl.
Note
follow the instructions for use for the amalgam separator.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.9 Preparation of the amalgam separator
26/32