▶ Top off the oxygenal (max. 1 litre)
Note
Do not fill through the centre opening (tumbler drain).Do not fill through the centre
opening (tumbler drain).
▶ Quit replenishing immediately once a signal sound is issued every second (10
sounds).
Reservoir is full.
▶ Attach the tumbler holder and the tumbler filler.
Note
Check the O-ring on the tumbler filler for damage.
3.1.1 Refilling oxygenal when using the water bottle
CAUTION
Damage caused by defective water bottle.
Malfunctions or failures from a leaky water bottle.
▶ Do not wash the water bottle in a dish washer, and do not rinse it with water
hotter than 50 ℃.
▶ Do not use the water bottle if it is scratched, deformed or discoloured.
▶ Note the maximum useful life of the water bottle on the label.
Water bottle complete: (Mat. no. 0.710.4151)
Care instructions ESTETICA E50
3 Replenish and check the disinfectant | 3.1 Replenish oxygenal