KaVo ESTETICA E50 - Checking the Oxygenal Concentration

KaVo ESTETICA E50
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Note
Substances remaining in the dosing head are not contamination, but rather agents
that are deposited there during extended periods of non-use after condensation of
the water. These deposits redissolve after several dosing processes.
3.2 Checking the oxygenal concentration
▶ Fill the tumbler with water.
▶ Use Merckoquant peroxide test strips 100 through 1000 mg/l H
2
O
2
(Merck article
number: 1100337) to determine the oxygenal concentration in the water disin‐
fection system.
The oxygenal concentration in normal operation should be between 150 and 250
mg/l H
2
O
2
.
Care instructions ESTETICA E50
3 Replenish and check the disinfectant | 3.2 Checking the oxygenal concentration
32/32