The selected rinse program starts.
or
▶ Turn off the device and do not restart it until the next morning.
The cycle is started automatically. The rinsing time follows rinsing program 02.
Note
If the device is not used the entire day (i.e., the instruments are not mounted), the
rinsing program automatically starts the next morning when the device is turned
on.
The rinsing programme can be stopped at any time by pressing the "intensive dis‐
infection" key.
If the device is turned off and then on again, the rinsing program also stops.
2.7.2 Intensive disinfection
Note
Observe the instructions for use for the instruments.
Intensive disinfection inhibits the formation of microorganisms when the equipment
is not used for extended periods of time. During intensive disinfection, the water
ducts are automatically filled with a higher concentration of oxygenal. The intensive
disinfection programme takes approx. 45 minutes and the increased hydrogen per‐
oxide concentration acts in the system for at least 30 minutes.
An additional intensive disinfection must be carried out during extended periods on
non-use (> 4 weeks).
The need for intensive disinfection is automatically displayed in the top LED of the
"Intensive disinfection" button at first start-up and after periods of non-use (week‐
end, holidays, vacation, etc.).
When a beep is issued every 10 seconds and the bottom LED of the "Intensive
disinfection" button flashes, Oxygenal must be replenished. If a beep is issued dur‐
ing the filling process (10 times each second), the container is full.
Note
The suction cannot be used while the HYDROclean step (4.6) is in progress.
Note
If the amalgam separator malfunctions during intensive disinfection, the entire
process is stopped. The bottom LED of the "HYDROclean" button flashes.
Intensive disinfection must be carried out:
▪ At first start-up
▪ After periods of non-use
▪ Over the weekend
▪ During vacations
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.7 Automatic hygiene function of the water-conducting
system
22/32