Start intensive disinfection
▶ Press the "Intensive disinfection" button until you hear a beep and the top LED
flashes. The display changes to the intensive disinfection menu.
▶ Prepare for intensive disinfection.
▶ Press the "Intensive disinfection" button.
Intensive disinfection starts and proceeds though several stages. The process
can be followed in the display. The major steps are listed below.
Step Description
7.0 Disinfection of the three- and multi-function syringe and spittoon bowl
60 seconds
6.0 Instrument disinfection
30 seconds each
5.0 Tumbler disinfection
40 seconds
4.0 Dwell time (signal sound, the process can be shut off)
30 minutes
3.0 Tumbler rinsing
40 seconds
2.0 Instrument rinsing
30 seconds
1.0 Rinsing the three- and multi-function syringe
30 seconds
0.1 Waiting position for setting down the instruments
Intensive disinfection is completed when "01 intensive disinfection completed" ap‐
pears on the display.
▶ Reinsert all instruments and remove the holder.
Interrupt intensive disinfection during the dwell time
So that no one will have to remain with the device throughout the entire intensive
disinfection process, intensive disinfection can be interrupted in programme step 4
(dwell time in which the microorganisms are killed by an increased oxygenal con‐
centration) (such as on the last workday before the weekend).
▶ Turn off the unit when the intensive disinfection program is in step 4 (an audible
signal sounds for 4 seconds).
When the device is switched on after 30 minutes of dwell time, the remaining pro‐
gramme steps run immediately.
If the device is turned on before the the 30 minutes are over, the dwell times finishes
first before the remaining steps of the intensive disinfection program run.