Skip exposure phase
This function is required when the disinfection procedure has already been started
and the treatment unit needs to be used (for example in an emergency, at hospitals,
etc.).
Intensive disinfection is in step 4:
▶ Press and hold down the "Intensive disinfection" button and the "Hydrocolloid"
button.
▶ Simultaneously press the foot pedal on the foot control.
The intensive disinfection switches to step 3.7.
The automatic rinsing process starts up.
Intensive disinfection during extended periods of non-use (> 4 weeks)
During extended periods of non-use, e.g. semester breaks at university, intensive
disinfection is carried out during the time of non-use. The disinfection fluid remains
inside the system during the period of non-use.
Before the period of non-use
▶ Start the intensive disinfection and stop it in step 4 after the signal sound during
the dwell time of 30 minutes.
▶ To prevent damage due to hoses getting kinked, remove hoses from the disin‐
fection attachment and place them in the dentist support.
▶ Keep the disinfection attachment in the spittoon bowl.
▶ Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶ Turn on the central compressor.
▶ Once the system pressure is established, turn on the central water supply for
the treatment unit.
▶ Replace the instrument hoses of the dentist and assistant units in the places
provided in the disinfection attachment.
▶ Turn the treatment unit on.
The unit automatically starts the remaining steps of the unfinished cycle.
▶ Allow the cycle to complete and then carry out a second full intensive disinfec‐
tion.
The unit is now ready for use.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.7 Automatic hygiene function of the water-conducting
system