▶ For the rinsing programme and intensive disinfection, press the seat adapter into
one of the four rubber seats of the disinfection attachment.
▶ Place the COMFORTbase in the seat adapter.
Carry out the rinsing programme
Set rinsing time
With the foot pedal of the foot control being pressed down, the rinsing time for the
respective rinsing programme can be set in the range from 0:01 to 4:00 minutes
using the "Decrease value" and "Increase value" buttons.
The rinse program proceeds in four steps. For rinse program 01, the cycle starts in
step 2, and for rinse program 02, the cycle starts in step 4.
Step Description
4.0 Fresh water supply
60 seconds
3.0 Tumbler rinsing
20 seconds
2.0 Instrument rinsing
The rinsing time can be set variably for each instrument.
Standard setting:
▪ Rinsing programme 01 = 20 seconds per instrument
▪ Rinsing programme 02 = 2 minutes per instrument
1.0 Rinsing the three- and multi-function syringe
set rinsing time of rinsing program 01 or 02
0.1 Waiting position for setting down the instruments
Start rinsing programme
Note
The rinsing program can be started immediately or the next morning.
▶ Briefly press the "intensive disinfection" key to immediately start the process.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.7 Automatic hygiene function of the water-conducting
system
21/32