Place the instruments on the sterilisation attachment
The cannulas of the three- and multi-function syringes must be taken off for the
rinsing programme and intensive disinfection.
▶ Remove the cannula with a slight amount of rotation while holding the tip on the
grip sleeve.
▶ Insert three- or multi-function syringes without cannulas in the tightening ele‐
ments of the disinfection attachment. Make sure that the "W" button (water) of
the three-function or multi-function syringes is pressed, and that no air is exiting.
▶ Place the instrument hoses of the dentist and assistant units in the provided
places of the disinfection attachment.
Note
All water regulation mechanisms for the motor, instruments, and multiflex couplings
must be open.
Place the COMFORTbase on the disinfection attachment
Note
A seat adapter (Mat. no. 1.005.6892) for the disinfection attachment is available
for the COMFORTbase.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.7 Automatic hygiene function of the water-conducting
system
20/32