Cleaning preparations Comply with product-specific reprocessing instructions.
Manual cleaning Fixtures:
Brush or cloth
Tap water 30 °C ± 5 °C
Rinse surface contamination off the product under running water.
Automatic cleaning
KaVo recommends thermodisinfection at up to 95 °C according to ISO
15883-1, e.g. Miele G 7781 / G 7881.
For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector.
Do not clean instruments and device components in the ultrasound
bath.
Manual disinfection Perform wipe disinfection only, the use of spray disinfection and dis‐
infection baths is not feasible.
Wipe all outer surfaces with a soft cloth and an approved disinfectant.
Approved disinfectants:
Microcide liquid (Schülke & Mayr)
FD 322 (Dürr)
Incidin Liquid (Ecolab)
CaviCide (Metrex)
Use disinfectants according to the manufacturer's Instructions for Use.
Please note the disinfectant safety data sheet.
Automatic disinfection
KaVo recommends thermodisinfection at up to 95 °C according to ISO
15883-1, e.g. Miele G 7781 / G 7881.
For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector.
Do not disinfect instruments and device components in the ultrasound
bath.
Manual drying Ensure dry, clean and uncontaminated compressed air according to
ISO 7494-2.
Blow off the outside and inside with compressed air until water drops
are no longer visible.
Automatic drying Automatic drying usually is part of the cleaning/disinfection cycle of the
thermodisinfector and should not be carried out above 95 °C.
Service Discard damaged products.
Control and function test Perform a visual inspection checking for damage and wear and tear.
Test products for function regularly.
Packaging Use standardised packaging material. Sealing should not be under
tension.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo
treatment units
4/32