Cleaning preparations ▶ Comply with product-specific reprocessing instructions.
Manual cleaning Fixtures:
▪ Brush or cloth
▪ Tap water 30 °C ± 5 °C
▶ Rinse surface contamination off the product under running water.
Automatic cleaning
KaVo recommends thermodisinfection at up to 95 °C according to ISO
15883-1, e.g. Miele G 7781 / G 7881.
▶ For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector.
▶ Do not clean instruments and device components in the ultrasound
bath.
Manual disinfection ▶ Perform wipe disinfection only, the use of spray disinfection and dis‐
infection baths is not feasible.
▶ Wipe all outer surfaces with a soft cloth and an approved disinfectant.
Approved disinfectants:
▪ Microcide liquid (Schülke & Mayr)
▪ FD 322 (Dürr)
▪ Incidin Liquid (Ecolab)
▪ CaviCide (Metrex)
▶ Use disinfectants according to the manufacturer's Instructions for Use.
▶ Please note the disinfectant safety data sheet.
Automatic disinfection
KaVo recommends thermodisinfection at up to 95 °C according to ISO
15883-1, e.g. Miele G 7781 / G 7881.
▶ For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector.
▶ Do not disinfect instruments and device components in the ultrasound
bath.
Manual drying ▶ Ensure dry, clean and uncontaminated compressed air according to
ISO 7494-2.
▶ Blow off the outside and inside with compressed air until water drops
are no longer visible.
Automatic drying Automatic drying usually is part of the cleaning/disinfection cycle of the
thermodisinfector and should not be carried out above 95 °C.
Service ▶ Discard damaged products.
Control and function test ▶ Perform a visual inspection checking for damage and wear and tear.
▶ Test products for function regularly.
Packaging ▶ Use standardised packaging material. Sealing should not be under
tension.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo
treatment units