Note
The return air filter is in the same position in the dentist unit variant with S table.
2.4.6 Reprocessing of the kit for physiological saline solution
The kit for physiological saline solution is optional.
Note
The silicone hoses of the kit must be sterilised after each patient.
Cleaning the kit for physiological saline solution
When using saline solution, all lines conducting the solution must be rinsed directly
after treating the patient.
▶ Pull out the suction hose from the NaCl bag.
▶ Immerse the pump hose needle in a container with distilled water (min. 150 ml).
▶ Activate the pump, and pump distilled water completely through it.
Sterilisation kit for physiological saline solution
Note
The following parts can be subjected to sterilisation:
– Silicone hoses for saline solution and their connecting nipples
CAUTION
Damage caused by sterilisation.
▶ Do not sterilise instrument hoses and hose clips for saline hose.
Requirement
The hoses for the physiological saline solution are rinsed.
▶ Seal the motor holder, handpieces, and contra-angle handpieces (and instru‐
ments if applicable) in a sterilisation bag and sterilise them.
Care instructions ESTETICA E50
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit