Page #17 background imageLoading...
Page #17 background image
Front panel tests
There are two front panel tests: one to test the functionality of the front panel keys and one to
test the display. In the event of a test failure, refer to “Display Board Checks” for details on trou-
bleshooting the display board.
KEY test
The KEY test allows you to check the functionality of each front panel key. Perform the fol-
lowing steps to run the KEY test:
1. Press SHIFT and then TEST to access the self-test options.
2. Use the
or
key to display “TEST: KEY”.
3. Press ENTER to start the test. When a key is pressed, the label name for that key is dis-
played to indicate that it is functioning properly. When the key is released, the message
“NO KEY PRESS” is displayed.
4. Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT
aborts the test and returns the instrument to normal operation.
DISP test
The display test allows you to verify that each pixel and annunciator in the vacuum fluores-
cent display is working properly. Perform the following steps to run the display test:
1. Press SHIFT and then TEST to access the self-test options.
2. Use the
or
key to display “TEST: DISP”.
3. Press ENTER to start the test. There are four parts to the display test. Each time ENTER
is pressed, the next part of the test sequence is selected. The four parts of the test se-
quence are as follows:
A. All annunciators are displayed.
B. The pixels of each digit are sequentially displayed.
C. The 12 digits (and annunciators) are sequentially displayed.
D. The annunciators located at either end of the display are sequentially displayed.
4. When finished, abort the display test by pressing EXIT. The instrument returns to
normal operation.
2-4 Troubleshooting

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD