34
Table 10. SENSe command summary
Command Description
[:SENSe[1]]
:FUNCtion <name>
:FUNCtion?
:DATA?
:FRESh?
:HOLD
:WINDow <NRf>
:WINDow?
:COUNt <NRf>
:COUNt?
:STATe <NRf>
:STATe?
Select measurement function:
‘VOLTage:AC’, VOLTage :DC’,
RESistance’, ‘FRESistance’, ‘CUR-
Rent:AC’, ‘CURRent: DC’ , ‘FRE-
Quency’,‘TEMPerature’, ‘PERiod’,
‘DIODe’, “CONTinuity’.
Query function.
Return the last instrument reading.
Returns a new (fresh) reading.
Path to control Hold feature:
Set Hold window (%); 0.01 to 20.
Query Hold window.
Set Hold count; 2 to 100.
Query Hold count.
Enable or disable Hold.
Query state of Hold.
:CURRent:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
Path to configure AC current.
Set integration rate (line cycles;
0.01 to 10).
Query line cycle integration rate.
Path to configure measurement
range:
Select range (0 to 3.1).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (-3.1 to 3.1).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution
(4 to 7).
Query resolution.
Path to configure and control the
filter.
Select filter type: (MOVing or
REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
2000-903-01A Page 34 Thursday, December 23, 1999 7:52 AM