System SourceMeter® Instrument Reference Manual Appendix B:
2600BS-901-01 Rev. B / May 2013 B-39
Step 8. Repeat calibration procedure for Model
2602B/2604B/2612B/2614B/2634B/2636B Channel B
For the Models 2602B/2604B/2612B/2614B/2634B/2636B only, repeat the entire procedure above for
Channel B. Be sure to:
• Make test connections to Channel B terminals.
• Substitute "smub" for "smua" in all commands.