Section
7: TSP command reference Series 2600B System SourceMeterĀ® Instrument
Reference Manual
7-206 2600BS-901-01 Rev. B / May 2013
smuX.measure.highcrangedelayfactor
This attribute contains a delay multiplier that is only used during range changes when the high-capacitance mode
is active.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Instrument reset
SMU reset
Recall setup
Saved setup 10
Usage
delayFactor = smuX.measure.highcrangedelayfactor
smuX.measure.highcrangedelayfactor = delayFactor
delayFactor
The delay factor; set to a value between 1 and 99
X
SMU channel (for example, smua.measure.highcrangedelayfactor applies
to SMU Channel A)
Details
This delay multiplier is only active when the high-capacitance mode is active.
Example
smua.measure.highcrangedelayfactor = 5
Increases the delay used during range changes for
SMU channel A by a factor of 5.
Also see
High-capacitance mode (on page 3-65)
reset() (on page 7-160)
setup.recall() (on page 7-180)
smuX.reset() (on page 7-216)
smuX.source.highc (on page 7-221)
smuX.measure.interval
This attribute sets the interval between multiple measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
0 (0 s)
Usage
interval = smuX.measure.interval
smuX.measure.interval = interval
interval
The interval value (in seconds); set to a value between 0 and 1
X
Source-measure unit (SMU) channel (for example,
smua.measure.interval applies to SMU channel A)