7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-348 2600BS-901-01 Rev. B / May 2013
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.timer[N].reset()
N
Trigger timer number (1 to 8)
Details
The trigger.timer[N].reset() function resets the following attributes to their factory defaults:
trigger.timer[N].count
trigger.timer[N].delay
trigger.timer[N].delaylist
trigger.timer[N].passthrough
trigger.timer[N].stimulus
It also clears trigger.timer[N].overrun.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 back
to factory default values.
Also see
trigger.timer[N].count (on page 7-344)
trigger.timer[N].delay (on page 7-345)
trigger.timer[N].delaylist (on page 7-345)
trigger.timer[N].overrun (on page 7-346)
trigger.timer[N].passthrough (on page 7-347)
trigger.timer[N].stimulus (on page 7-348)
trigger.timer[N].stimulus
This attribute specifies which event starts the timer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Not saved
0
Usage
eventID = trigger.timer[N].stimulus
trigger.timer[N].stimulus = eventID
eventID
The event that triggers the timer delay
N
Trigger timer number (1 to 8)