This command copies the script from nonvolatile memory into the run-time environment. It also creates a global
variable with the same name as the name of the script.
Example
script.restore("test9")
Restores a script named "test9" from nonvolatile
memory.
Also see
script.delete() (on page 7-163)
script.run()
This function runs the anonymous script.
Type TSP-Link accessible
Affected by Where saved Default value
Function No
Usage
script.run()
run()
Details
Each time the script.run() command is given, the anonymous script is executed. This script can be run
using this command many times without having to re-send it.
Example
run()
Runs the anonymous script.