EasyManuals Logo
Home>Keithley>Measuring Instruments>2601B

Keithley 2601B Reference Manual

Keithley 2601B
864 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #370 background imageLoading...
Page #370 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-30 2600BS-901-01 Rev. B / May 2013
Example 2
printbuffer(1, 6,
smua.nvbuffer1.sourcevalues)
Print the sourced value of the first 6 readings
stored in source-measure unit (SMU) A,
buffer 1.
Example output:
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04
Also see
bufferVar.measurefunctions (on page 7-22)
bufferVar.measureranges (on page 7-23)
bufferVar.n (on page 7-24)
bufferVar.readings (on page 7-25)
bufferVar.sourcefunctions (on page 7-26)
bufferVar.sourceoutputstates (on page 7-27)
bufferVar.sourceranges (on page 7-28)
bufferVar.statuses (on page 7-30)
bufferVar.timestamps (on page 7-32)
Reading buffers (on page 3-6)
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Clearing the buffer
See Details
Not applicable
Usage
statusInformation = bufferVar.statuses[N]
statusInformation
The status value when reading N of the specified buffer was acquired
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
N
The reading number (1 to
bufferVar
.n)

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals