9: Integrating with a Model 3706A-S using TSP-Link Model DMM7510 7½ Digit Multimeter
9-8 DMM7510-900-01 Rev. B / May 2015
node[2].trigger.model.setblock(16, node[2].trigger.BLOCK_CONFIG_NEXT,
"myScanConfigList")
node[2].trigger.model.setblock(17, node[2].trigger.BLOCK_WAIT,
node[2].trigger.EVENT_TSPLINK1)
node[2].trigger.model.setblock(18, node[2].trigger.BLOCK_MEASURE)
node[2].trigger.model.setblock(19, node[2].trigger.BLOCK_NOTIFY,
node[2].trigger.EVENT_NOTIFY1)
node[2].trigger.model.setblock(20, node[2].trigger.BLOCK_BRANCH_COUNTER,
muxChannelsCount[4], 17)
--Initiate trigger models on both Model 3706A-S and Model DMM7510 and wait until
finished
node[2].trigger.model.initiate()
-- To guarantee that the Model DMM7510 trigger model is at WAIT block (block 2)
delay(2)
node[1].scan.background()
waitcomplete()
--Open all channels on slot 1 and 4
node[1].channel.open("slot1, slot4")