Tab. 31: Error codes
7.5 Warning codes digital LED display
Code
Cause
Note
SEr
(A01)
Service due
Perform service
CLE
(A02)
Differential pressure too high
Carry out cleaning (see
“Maintenance” chapter)
A02
Vacuum in extraction pipe
too high
Check the extraction pipe for
blockage
A03
Extraction volume
measurement sensor error
The digital LED display shows the
motor frequency, contact service
A05
Extraction capacity not
sufficient.
Horn active
The required extraction
capacity for the connected
extraction torch is not
reached.
Check the extraction torch for
blockage.
Carry out a filter change.
A06
Control temperature
increased (15 ⁰C below
switch-off limit)
No measure required
A07
Temperature in the unit
increased (15 ⁰C below
switch-off limit)
No measure required
Tab. 32: Warning codes
NOTE
Overload protection
If the vacuum in the extraction pipe rises above 22,000 Pa and the
temperature in the product exceeds a certain threshold, the product
automatically reduces the extraction capacity to avoid overloading the
product.
NOTE
If the fault cannot be corrected by the customer, please contact the
manufacturer’s service department.