Before applying power to your device make sure that the protective connector of the 3
conductor mains power cord is correctly connected to the protective earth contact of the
socket outlet.
Never use a mains power not supplied by KEOPSYS.
IMPROPER GROUDING CAN CAUSE ELECTRIC SHOCK with damage to your health
and even death.
Never use a unit without its complete package including all covers ; ask if necessary for
a replacement part.
The disconnection of the equipment from the mains can be performed by the
disconnection of the AC input power cord connected to the equipment..
1.2.6. Other safety considerations
The KEOPSYS products comply with the European and International normalization EMC IEC 61326.
The device has been designed according the standard relative to “Safety requirements
for electrical equipment for measurement, control and laboratory use”
Do not use the device out of this type of application.
The device is designed for an inside application.
The device must be properly located on a flat horizontal surface.
The device must be located in order to keep free the fan air outlets at the rear of the
box.
The environmental conditions during the use of the device must comply with those
described in the technical specification.
1.2.7. Environmental conditions
• Equipment dedicated to measurement, control and laboratory use only.
• Voltage range: 100-240V +/-10%.
• Maximum input current is limited to 3.15A.
• The presumed short-circuit current of the mains power connection interface shall be limited to
1500A.
• The Overvoltage Category : Cat II
• Temperature rating : (5°C to 40°C)
• Maximum altitude: 2000m.
• Relative Humidity range: 0% to 80% until 31°C with linear decrease down to 50% until 40°C.
• Pollution degree : 2
• Maximum input current is limited to 3.15A, the presumed short-circuit current of the mains
power connection interface shall be limited to 35A.