EasyManua.ls Logo

KERN CDS 4K0.02 - External Adjustment with User-Defined Adjustment Weight < Caleud

KERN CDS 4K0.02
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TCDS-BA-e-2210 18
Place the adjustment weight and confirm with -button,
< wait > followed by < reMvld > will be displayed.
Once < reMvld > is displayed, remove the adjustment
weight.
After successful adjustment the balance automatically
returns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show the error message < wrong>.
Switch off balance and repeat the adjustment process.
7.8.2 External adjustment with user-defined adjustment weight < caleud >
Press and hold the TARE and ON/OFF buttons
simultaneously to enter the setup menu.
Wait until the first menu item < Cal > is displayed.
Confirm by button, < CalExt > will be displayed.
Use the navigation keys to select < caleud >.
Acknowledge by -button. The numeric input window for
the weight value of the adjustment weight appears. The
active digit is flashing.
Provide adjustment weight.
Enter weight value, numeric input see chap. 3.2.2