TCDS-BA-e-2210 42
Buttons
Allocate function
key
Not documented
Blight
Display
background
illumination
mode
always
Background lighting of display is
switched on permanently
timer
The background illumination is
automatically switched-off according
to the time without load change or
without operation defined in menu
item < Time >
No bl
Display background illumination
always switched off
Time
5 s
Definition, after which time the
background illumination is
automatically switched-off without
load change or without operation.
10 s
30 s
1 min
2 min
5 min
30 min
tarerg
Taring range
100%
10%
Definition max. taring range, selectable 10% - 100%.
Numerical input, see chap. 3.2.2.
ztrack
Zerotracking
on
Automatic zero tracking [ <3d ]
off
In the event that small quantities are removed or added
to the material to be weighed, incorrect weighing results
can be displayed due to the “stability compensation“.
(e.g. slow flow of liquids from a container placed on the
balance, evaporating processes).
When apportioning involves small variations of weight, it
is advisable to switch off this function.
reset
Reset balance settings to factory settings

Table of Contents

Other manuals for KERN CDS 4K0.02

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CDS 4K0.02 and is the answer not in the manual?

KERN CDS 4K0.02 Specifications

General IconGeneral
BrandKERN
ModelCDS 4K0.02
CategoryScales
LanguageEnglish

Related product manuals