CFS/CCS-BA-e-1624 59
(2) Adjustment of bulk scales, model CFS 50K-3
1. Internal resolution
Press , the internal resolution will be shown.
Return to menu by .
Select the next menu point „dESC“ with .
„Cnt“
2. Position decimal point /
Press on display „dESC“; the currently adjusted position
of the decimal point will be shown.
Select desired setting with and acknowledge by .
Use to select the next menu item [CAP].
„dESC“ „0.00“
CAP
Press on display „CAP“; the currently adjusted capacity
will be shown.
Select desired setting with and acknowledge by .
Delete with for changes, and input desired value with
the numeric keys.
Confirm input with , the balance returns to the menu.
Select the next menu point „div“ with .
„CAP“
„SEL“ „060.000“
„CAP“
Press , the current setting will be displayed.
Select desired setting by and confirm with , the
balance returns to the menu.
Select the next menu point „AZt“ with
„inC“ „5“