56 KXS_KXG-TM-BA_IA-e-1932
Should it be sent continuously, the command must be preceded by %.
If stable weight values are to be sent, the command must be preceded by #.
Command Description (RJ, RK, RL, RS)
RJ
RK
RL
Hi/ Lo/ OK status Sample display
Lo (001) 001+ 2.000
OK (010) 010+ 3.000
Hi (100) 100+ 4.000
RS
Show tolerance limit
RS
: Class (00 ~ 99) *  : LO or HI
HI Display of the adjusted upper limit value
LO Display adjusted lower limit value
Example: Command
RS02LO<CR><LF>
Response RS02LOXXXXXX<CR><LF>
* depending on model
13.5.3 Command Format C
Host
Command+ Data
Slave
Command+ Data
WS
Set tolerance limit
WS XXXXXX

: Class (00 ~ 99) *

: LO or HI XXXXXX: Enter the limit value
HI Enter the value for the upper limit
LO Enter the value for the lower limit
Example: Command
WS00HI001000<CR><LF>
Response WS00HI001000<CR><LF>
* depending on model