KXS_KXG-TM-BA_IA-e-1932 65
4. Gravitation constant at place of installation
5. Gravitation constant applied during verification
Press repeatedly to exit menu.
15.4.1 Configuration example single range scale
Invoke menu item<mult> (see chap. 15.4) and press to
confirm. The currently set balance type is displayed.
Press to select desired type
SinGLE = single range scale
Press , the screen used to enter readability/verification
value will be shown.
Press , the current setting will be displayed.
Select desired setting with and acknowledge by .
Press to select next menu item for entering capacity.
Press , the current setting will be displayed.
Press to select the desired setting.
Confirm by , „End“ will be shown.
To import the configuration data, press again.
The entering of configuration data requires subsequent
adjustment or linearization.
For how to carry out adjustment see chap. 7.7 or
linearization see chap. 7.8