To confirm, press and the current setting will be shown.
noLin = Adjustment
LineAr = Linearisation
For adjustment press to select setting < noLin> and confirm by .
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press .
The currently set adjustment weight will be displayed.
Either apply the displayed adjustment weight or make changes with the help of
the navigation keys (see chap. 3.1). Confirm by , „LoAd“ will be shown.
Carefully place adjustment weight in the centre of the weighing plate.
Wait for stability display, then press .
After the adjustment the balance will carry out a self-test. Remove adjustment
weight during selftest, balance will return into weighing mode automatically.