KN-4006 series UV Radiation Treatment System
Guidance and manufacturer's declaration – electromagnetic immunity –for all EQUIPMENT and
SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The KN-4000 series UV Radiation Treatment System is intended for use in the electromagnetic environment specified below.
The customer or the user of The KN-4000 series UV Radiation Treatment System should assure that it is used in such an
environment.
Electromagnetic environment-guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrostatic
transient / burst
EN 61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
± 2 kV for power supply lines
± 1 kV for input/output lines
Mains power quality should be that of a
typical commercial or hospital environment.
±1 kV differential mode
±2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a
typical commercial or hospital environment.