KN-4006 series UV Radiation Treatment System
70
continued table
Immunity test
EN 60601 test level
Compliance level
Electromagnetic environment-guidance
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN 61000-4-11
< 5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
< 5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of The KN-4000 series UV
Radiation Treatment System requires
continued operation during power mains
interruptions, it is recommended that The
KN-4000 series UV Radiation Treatment
System be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NOTE: U
T
is the a. c. mains voltage prior to application of the test level.