KN-4006 series UV Radiation Treatment System
71
Guidance and manufacturer´s declaration electromagnetic immunity for EQUIPMENT and SYSTEM
that are not LIFE-SUPPORTING
Guidance and manufacturer´s declaration electromagnetic immunity
The KN-4000 series UV Radiation Treatment System is intended for use in the electromagnetic environment specified
below. The customer or the user of The KN-4000 series UV Radiation Treatment System should assure that it is used in such
an environment.
Immunity test
EN 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of The KN-4000
series UV Radiation Treatment System, including cables,
than the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
P
V
d ]
5.3
[
1
=
P
E
d ]
5.3
[
1
=
P
E
d ]
7
[
1
=