Keysight 16452A Liquid Test Fixture
Operation and Service Manual
3 Operation
Introduction
This chapter describes how to measure the dielectric constant () of the liquid
using the 16452A.
The standard measurement sequence is shown in this chapter.
Some instructions for the temperature measurement are provided at the
back of the chapter.
Measurement Sequence
Relative permittivity (
r
) can be calculated from the ratio of the capacitance of
a material to that of air (nearly equal to that of a vacuum). The standard
measurement sequence is as follows:
1. Prepare the test fixture for use.
2. Connect the test fixture.
3. Check the SHORT residual.
4. Set the instrument for capacitance measurement.
5. Do a SHORT compensation.
6. Do the air capacitance (C
o
) measurement.
7. Do the liquid capacitance and resistance (C
p
, R
p
) measurement.
8. Drain the liquid (after measurement)
9. Do the data processing- Calculate dielectric parameter from measurement
data.