42 Keysight 16452A Liquid Test Fixture
Service
Assembly Replacement
4-
Figure 4-3 Inside of the Electrode Assembly
Assembling Procedure for Electrode
See Figure 4-2 and Table 4-3 for the reference designators.
1. Pass the SMA cable assembly through the insulator and the O ring
.
2. Insert the cable into the electrode body, and connect it using the insulators
, the washers and the screws .
3. Bend the semi-rigid cable at 33 mm from the top face of the electrode body
as described in Figure 4-3, and set the angle between the semi-rigid cables
close to 90º.
4. Solder the semi-rigid cables to the center pin of the electrode.
5 2190-0654 Washer, M2 8 16452-60001/
16452-60002
High Electrode
Assembly/
Low Electrode
Assembly
1
6 0515-0976 Screw, M2 8 16452-60001/
16452-60002
High Electrode
Assembly/
Low Electrode
Assembly
1
Table 4-3 16452A Replaceable Parts (Around SMA Connector)
Ref
/D
Part Number Description Qty RoHS
Compliant
Replacement
Part
Description Qty